Integrated Biobanking Workflows: Barcode scanning failure on automated nucleic acid extractor

September 30th, 2015 in Working Group Update

INTEGRATED BIOBANKING WORKFLOWS WORKING GROUP:
Workflow Case Studies and Lessons Learned Series

Workflow Case Study 4 – Barcode scanning failure on automated nucleic acid extractor

Contributions by: Conny Mathay, Gaël Hamot, and Michaël Heymann (Integrated Biobank of Luxembourg); and Angela Notah-Chavez (ISBER News Assistant Editor)


Introduction

This case study is reviewing the difficulties with scanning biospecimen sample 2D barcodes in an automated nucleic acid extractor workflow.  Scanning failures may occur due to issues with sample ID format, thereby bringing to a halt the associated liquid handler automated DNA extraction run.

IMG_0430

IMG_0429


Problems

  • Barcode is encoded with a sample ID which is incompatible with the device’s software as the sample ID is alphanumeric and contains too many characters.
  • There is a difference between the maximum number of alphanumeric characters in the LIMS versus the liquid handler software.

Because the data cannot be processed (buffer overload), the extraction run stops.  The samples must then be extracted via the extractor standalone protocol.


Workflow Background

SOPs

  • Automated DNA Extraction from Non-fixed Materials
  • Use and Maintenance: Nucleic Acid Extractor with Automated Liquid Handling Workstation

Equipment

  • Nucleic acid extractor with automated liquid handling workstation

Consumables

  • Adhesive labels
  • DNA blood extraction kit for 24-rod head
  • Falcon tubes, 4mL

IMG_0427


Findings/Observations

The issues with the scanning failures may be time-consuming, particularly when switching from the automated extraction protocol to the standalone extraction protocol.  More significantly, if an automated DNA extraction run stops, the integrity of the samples may be compromised as they will be sitting at room temperature.


Solution

Change the script and software settings so that a sample ID of up to 70 alphanumeric characters may be accepted.

– – – – – – – – – – –

Call for participation!

The Integrated Biobanking Workflows Working Group is recruiting members to develop more case studies based on this same template and where the objective is to uncover points in workflow integration which require improvement.  Case studies may come from either automated or manual processes, from processes at any throughput level, and from a biorepository of any type and size.